Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: Triple Quad LC-MS
  • Manufacturer: Varian
  • Country: USA
  • Available: Discontinued
All specifications

About

Short description

Retired triple quadrupole LC/MS system; Electrospray/APCI ionization; Wide scan and MRM modes

Specifications

Retired triple quadrupole LC/MS system; Electrospray/APCI ionization; Wide scan and MRM modes

  • Status: Discontinued
  • Country: USA
  • Analyzer type : Tandem Quadrupole MS
  • Scan modes : Full scan, SIM, Precursor scan, Product scan, Neutral loss scan, SRM, MRM
  • Mass range : 10 to 2,000 Da
  • High resolution : No
  • Mass resolution/ Mass resolving power : Unit mass resolution
  • Ion source types (LC-MS) : ESI (standard), APCI (optional)
  • Dynamic range : 5 orders of magnitude
  • Sensitivity : +ESI SIM: 5 pg Reserpine ≥10:1; +ESI MS/MS: 5 pg ≥20:1
  • Mass accuracy/Stability : ±0.1 Da over 24 h
  • Scanning speed : 500-6000 Da/sec
  • Dwell time : <5 ms adjustable
  • MRM transition rate : -
  • Polarity switching : 240 ms
  • Fragmentation : CID via 180° curved collision cell
  • Vacuum system : Dual-stage turbomolecular pump (310/400 L/sec) + 1 foreline pump
  • Power Requirements and Consumption : MS: 100-240 Vac, 50/60 Hz ±3 Hz, 1.0 kW; Foreline: 100-240 Vac, 50/60 Hz, 2.8 kW
  • Environmental requirements : Humidity: 20-80% RH (non-condensing); Temp: 16-30 °C
  • Dimensions and Weight: 50.0 × 38.0 × 71.0 cm; 108.0 kg