Specifications

  • Category: Liquid Chromatography - Mass Spectrometry
  • Technique: Single Quad LC-MS
  • Manufacturer: Microsaic Systems
  • Country: UK
  • Available: Active
All specifications

About

Short description

Compact single quadrupole MS. Integrated vacuum and PC. Point-of-need detection.

Specifications

Compact single quadrupole MS. Integrated vacuum and PC. Point-of-need detection.

  • Status: Active
  • Country: UK
  • Analyzer type : Quadrupole MS (single)
  • Scan modes : Full scan, SIM, simultaneous scan/SIM, timed SIM
  • Mass range : m/z 50 to 1,400
  • High resolution : No
  • Mass resolution/ Mass resolving power : 0.7 m/z ± 0.1 at FWHM* (*20 ± 3 °C)
  • Ion source types (LC-MS) : ESI (spraychip electrospray ionization source)
  • Dynamic range : 3-4 orders of magnitude
  • Sensitivity : 10 pg reserpine yields a peak in SIM mode with S/N = 10:1 (RMS)
  • Mass accuracy/Stability : ±0.3 m/z in full scan* (*20 ± 3 °C)
  • Scanning speed : -
  • Dwell time : -
  • Polarity switching : Yes
  • Fragmentation : -
  • Vacuum system : Integrated oil-free pumps (no floor pump needed)
  • Power Requirements and Consumption : -
  • Environmental requirements : 20 ± 3 °C for accuracy and resolution specs
  • Dimensions and Weight: 55.0 × 35.0 × 25.0 cm; 32.0 kg including PC, pump, exhaust